K-space Optimized Spatial Temporal Architecture for Low Latency in Interventional Cardiac MRI

M. Buckup, N. Mahajan, A. E. Rodriguez-Soto, J. Blood, S. Hegde, E. L. Schuchardt, B. Printz, H. K. Narayan, B. L. Gordon, F. Contijoch

PreprintUso en el mundo real

En palabras de los autores

Purpose: To develop and evaluate a recurrent neural network that operates directly on undersampled golden-angle radial k-space, enabling low-latency cardiac cine reconstruction for interventional cardiac MRI (iCMR). Methods: KOSTAL (K-space Optimized Spatial Temporal Architecture for Low latency) combines a per-coil k-space convolutional gated recurrent unit (convGRU), which learns a mask-conditioned view-sharing rule over the accumulated k-space, with an image-space recurrent module. It was trained and tested on the open-access multi-coil OCMR dataset using 10 simulated golden-angle spokes per frame against a 1000-spoke reference. Ablations varied recurrent module architecture and placement; temporal behavior was assessed using a imaging-plane transition test. Generalization was evaluated without retraining in 23 pediatric patients prospectively scanned during free-breathing using a radial golden-angle bSSFP sequence with binned XD-GRASP and clinical breath-held cine serving as references. Results: At 10 spokes per frame, KOSTAL reached a median SSIM of 0.907 (gridding: 0.622) in under 34 ms of processing time per frame. Image quality (IQ) matched gridded results formed from 7-11 times more spokes. Recurrence in k-space rather than image space accounted for most of this gain (0.891 versus [≤]0.650). IQ recovered from an abrupt plane transition in 75 ms, versus 303 ms for 200-spoke gridding. KOSTAL had the best agreement with the binned reference and the highest long-axis CNR. KOSTAL accurately tracked left-ventricular blood-pool volume (end-diastolic volume bias -0.7 mL) with a shorter lag (71 ms) than iterative reconstruction (239-287 ms). Conclusion: Learned, mask-conditioned recurrence in k-space yields high image quality at a latency compatible with interventional guidance.

Resultado principalEl resumen no menciona limitaciones.

Apareció: lunes, 28 de septiembre. medRxiv. Preprint, todavía sin revisión por pares.

DOI: 10.64898/2026.09.21.26363257