Design and Performance Evaluation of a Quad-panel Dual-isotope PET Imaging System

Yu Shi, Tengjiao Wang, Sen Yang, Bo Wen, Meng Niu, Yirong Wang, Risheng Yang, Yanyun Liu, Wei Ding, Xiangyu Liu, Fei Kang, Shouping Zhu

PreprintUso en el mundo real

En palabras de los autores

In simultaneous dual-isotope PET, prompt-gamma-based separation is challenged by the low statistics of triple coincidences and the limited flexibility of coincidence logic for multi-energy event extraction. We designed and evaluated a compact quad-panel dual-isotope PET system without dedicated prompt gamma detectors. The system comprised four orthogonally arranged detector panels, each providing an effective area of 107.0 * 145.5 mm. GPU-accelerated software coincidence processing extracted double- and prompt-gamma-tagged triple coincidences from raw single-photon data. Triple-coincidence images were reconstructed with correlated-random-triple correction using the consistent delayed expectation-maximization method, followed by image-domain isotope separation using a spatially varying scaling factor. System performance was evaluated using phantom and in vivo studies, with emphasis on 68Ga/18F imaging because of the low prompt gamma yield. The system achieved energy resolutions of 15.7%, 12.9% and 10.2% at 511, 1077, and 1275 keV, and a timing resolution of 1.35 ns. In the 68Ga Derenzo, 1.6-mm rods were resolved in double-coincidence images, with 1.2-mm rods partially distinguishable, while 1.6-mm rods remained discernible in triple-coincidence images. The mean prompt gamma detection efficiencies along the three axes ranged from 8.19% to 9.31%. The system sensitivity was 1.12% for doubles with an energy window of 350-650 keV, and 0.068% for triples after random correction with a dual-energy window. The scatter fraction was 27.81%, and the NECR reached 83 kcps at 9.38 MBq for doubles and 106 cps at 21.42 MBq for triples. Phantom and in vivo studies demonstrated successful separation of 18F/68Ga and 18F/124I tracer distributions. By maintaining high-resolution and quantitative imaging performance, the system provides a preclinical platform for multitracer PET studies.

Resultado principalEl resumen no menciona limitaciones.

Apareció: martes, 10 de noviembre. arXiv. Preprint, todavía sin revisión por pares.