EIT-based non-invasive measurements correlate with pulmonary arterial pressure in patients
In the authors' words
Abstract Objective. To investigate the relationship between cardiac synchronized timing of electrical impedance tomography (EIT) signals and invasively measured mean pulmonary artery pressure (mPAP) in patients with cardiovascular disease. Approach. EIT, ECG and mPAP were recorded in seven patients as part of routine surgery during an apnoea phase, while pressure changes were induced with phenylephrine. Four waveform timings were considered: the maximum, the maxima of the first and second derivatives, and the tangent intersection. Their delays from ECG R waves defined eight markers across the heart and lung regions. Main results. Across 56 patient marker combinations, the mean absolute Spearman coefficient was 0.74. Marker strength and direction varied across patients and landmarks. Each patient had at least one timing marker with an absolute value above 0.79. The lung second derivative timing marker exceeded 0.90 in four patients, more often than any other marker, but showed weaker associations in the remaining three. Significance. These findings represent the first study of EIT timing and mPAP in patients; it suggests that EIT could be a useful non invasive predictor of clinical pulmonary haemodynamics. However, work is required to combine information from timing markers and to relate EIT values to pressure.
In plain words AI summary
Researchers measured electrical impedance tomography (EIT), ECG, and mean pulmonary artery pressure in seven patients during surgery. They tracked eight timing markers based on delays from ECG R waves to EIT signal features in heart and lung regions. Across all patient-marker combinations, the average correlation with pressure was 0.74. Each patient had at least one marker correlating above 0.79 with pressure. The study suggests EIT timing could help predict pulmonary blood pressure without invasive procedures, but more work is needed to combine markers.
Written by Kimi K2.5 using only the abstract above. Jev checked that every sentence is supported by the abstract sentences it cites, and removed any that were not. Point at a sentence to highlight its source.
Appeared: Monday, November 2. medRxiv. Preprint, not yet peer-reviewed.