Ultrasound neural interface enabling acute spinal cord monitoring and blood flow modulation

Max Kerensky, Siddharth Krishnan, Yuxin Chen, Kelley M. Kempski Leadingham, Victor M. Quiroz, Joshua Punnoose, Constantin Smit, Denis Routkevitch, Avisha Kumar, A. Daniel Davidar, Leonard Steger, Carly Weber-Levine, Kelly Jiang, Meghana Bhimreddy, Stuart J. Bauer, Angelica F. Lopez, Patrick J. Kramer, Kayla Robinson, Arjun Kumar Menta, Abdul Karim Ghaith, Andrew M. Hersh, Haley G. Abramson, Ali Mohammadabadi, John N. Theodore, Osamah M. Zmily, Ritvik Jillala, Vikas N. Vattipally, David Kopala, Siavash Shirzadeh Barough, Mohammad Hossein Amini, Jonathan S. Maltz, Francisco Chavez, Kyle Morrison, Laura Roy, Ian Suk, Ryan Patrick Lee, Betty Tyler, Mark Luciano, Nitish V. Thakor, Joshua C. Doloff, Nicholas Theodore, Amir Manbachi

Peer-reviewed journalReal-world use

In the authors' words

This journal does not let us republish the full abstract. Here are the two sentences Pipette selected, quoted from it. Read the rest at the publisher.

Main result
We present the Multimodal Ultrasound System for Injury Care (MUSIC), a temporary epidural ultrasound neural interface placed during standard decompression surgery to enable postoperative brightness-mode imaging, Doppler-derived hemodynamic monitoring, and future device-integrated ultrasound actuation.
Limitation the authors admit
Across simulations, benchtop, human cadaveric, and porcine studies, MUSIC enabled on-demand visualization of cord structure, hematoma tracking, detection and longitudinal tracking of localized Doppler flow signatures attributable to the sulcal artery region, and, in healthy pigs, proof-of-concept focused ultrasound modulation of blood-flow velocity.

Appeared: Friday, November 6. Science Advances. Peer-reviewed journal.

DOI: 10.1126/sciadv.aeg2342