Laboratory Characterization of a Ge-CCD Prototype for Future Near-Infrared Astronomical Instrumentation

Emi Tamura, Juan Cruz Estrada Vigil, Prashansa Mukim, Giovanni Pinaroli, Christopher Leitz, Ilya Prigozhin, Kevan Donlon, Santiago Ezequiel Perez, Sven C. Herrmann

Preprint

In the authors' words

Germanium charge-coupled devices (Ge-CCDs) extend CCD imaging beyond the silicon cutoff and are candidate detectors for the near-infrared channel of future astronomical instruments, in particular the spectrographs of next-generation wide-field surveys. We report the laboratory characterization at Brookhaven National Laboratory (BNL) of a prototype fabricated at MIT Lincoln Laboratory. We describe the cryogenic test system and readout, and present measurements of the output waveform, dark current, readout noise, and X-ray response. The dark current at 94.69 K is e pixel s. Under Fe irradiation without on-chip binning, the charge generated by an X-ray is confined within pixels, and the Mn K line has a full width at half maximum (FWHM) of % for single-pixel events. Since the charge generated by X-rays above about 30 keV spreads over or more pixels, the CCD was operated with on-chip binning at higher energies. This yielded FWHMs of % at 30.9 keV (Ba, ASCA grades 0, 2, 3, 4, and 6) and % at 59.5 keV (Am, all grades). These results establish a baseline for the device and for the BNL test system. Measurements of the temperature dependence, optical response, and charge-transfer performance will connect this baseline to the requirements of near-infrared astronomical instruments.

Main resultLimitation the authors admit

Appeared: Monday, October 19. arXiv. Preprint, not yet peer-reviewed.